You are in: Instruments > AcqCam > Performance and Use > AcqCam Components > Detector |
![]() |
Acquisition Camera Detector |
The Acquisition Camera in both telescopes employs a 1024 x 1024 EEV CCD (CCD47). The pixel size is 13 microns (~0.12"). The CCD can be operated over a range of temperatures controlled through an interface to a thermoelectric cooler. The device full well is just over 70,000 counts/pixel. The CCD employs a 16 bit encoder which saturates at 65,536 counts, i.e. somewhat before the CCD full well. The minimum exposure time is 0.006 s. The full frame readout time is about 1.7 seconds.
Gemini South
At a temperature of -30 oC the bias level is ~550 ADU and
the dark current is ~ 13 electrons/pixel/second. The gain setting
is 1.0 (left amplifier) and 1.03 (right amplifier) electrons/ADU. The read
noise on the telescope is ~5 electrons rms (in the HIA lab it was 4.4 and
4.6 electrons rms for the left and right channel, respectively).
Laboratory measurements at HIA also indicated good charge transfer efficiency
(CTE), with 1-CTE = 5.22E-6 (parallel) and 3.18E-6 (serial). Departure
from linearity will be determined from data taken during commissioning
(in the HIA lab, the maximum departure from linearity was ~7%).
Pixels | 1024 x 1024 | 13 microns/pix | 0.12 arcsec/pix |
Gain | 1.0 (left amp) | 1.03(right amp) | |
Dark Current | ~12 e-/pix/sec | ~ 12 ADU/pix/sec | @T = -30 oC |
Read Noise | ~ 5 e- rms | ~ 5 ADU rms | |
Full Well | 70,000 ADU | ||
ADC Saturation | 65,536 ADU | ||
Linearity | (to be characterised) |
Gemini North
At a temperature of -30 oC the bias level is ~260 counts
and the dark current is ~9.5 counts/pixel/second. It maybe necessary to
run the detector at a warmer temperature to avoid condensation on the vacuum
window. This will result in a higher dark current. The gain setting is
0.74 (left amplifier) and 0.71 (right amplifier) electrons/ADU. The read
noise on the telescope is ~6 electrons rms. (In the HIA lab it was 4.4
and 4.6 electrons rms for the left and right channel, respectively.)
The maximum difference from linearity measured in the HIA lab was 6.9%.
Laboratory measurements at HIA also indicated good chargmeasurements at
HIA also indicated good charge transfer efficiency (CTE), with 1-CTE =
4.77E-7 (parallel) and 2.32E-5 (serial). Because the CCD always operates
with both amplifiers, the expected signal loss in the serial register corresponds
to a 1% edge-to-center photometric drop off for point sources.
Pixels | 1024 x 1024 | 13 microns/pix | 0.12 arcsec/pix |
Gain | 0.74 (left amp) | 0.71 (right amp) | |
Dark Current | ~7 e-/pix/sec | ~9.5 cnts/pix/sec | @T = -30 oC |
Read Noise | ~6 e- rms | ~8 cnts rms | |
Full Well | 98,940 e- | 70,000 cnts | 70,000 cnts |
ADC Saturation | 93,620 e- | 65,536 cnts | |
Linearity | < 3% non-linearity (to be characterised) |
Last update February 14, 2002; Cláudia Winge
In original form June 3, 2000; Ted von Hippel