| You are in: Instruments > GNIRS > Performance and Use > GNIRS Components > Gratings > Usage: long camera & 32 l/mm |
|
GNIRS Grating Usage: long camera and 32 l/mm |
| Resolving power: (lambda / delta lambda) |
5100 with 0.1 arcsec (2 pixel) slit |
| Typical uses: (...and limitations) |
(1) Spectrum of the
M-band (5um window) with 49 arcsec long slit
The wavelength coverage is about 0.45um so two grating settings are required to cover the entire window. With a wider slit to reduce slit losses the resolving power is degraded e.g. 0.3 arcsec slit yields R=1700 for an object that fills the slit. One can observe in the M-band with the short camera and the 110 l/mm grating, the maximum exposure time is ~0.5sec with the 0.3arcsec slit; if the objects are sufficiently bright that slit losses are not an issue, use of the long camera allows ~3x longer exposure times (thereby reducing overhead). Example spectra shown below of a standard star: 4.4 - 5.1um in 1st order
|
| (2) Intermediate
resolution spectra at any wavelength with fine spatial sampling
This mode gives approximately the same resolution as the
short camera with the 110 l/mm grating, using a
comparable slit width in pixels, but with 3x the spatial resolution. Because of the significant
slit losses
(e.g., a 0.1 arcsec slit is required to match the resolution of the 0.3arcsec
slit with the short camera), this mode is only recommended when very
high spatial resolution is desired. |
|
| Example spectra and other figures: | This mode is still in commissioning; additional sample spectra will be provided when available. |
| Fig 1: long-slit spectrum in 1st order (M-window, central wavelength 4.7um) |
![]() |
Last update August 24, 2004; B. Rodgers