GMOS-N Array
The GMOS detector array consists of three 2048x4608 EEV chips arranged in a row. The table below gives a summary of the current detector/controller characteristics.
| Array | EEV | ||
| Pixel format | 6144x4608 pixels | ||
| Array layout | Three 2048x4608 chips in a row with ~0.5mm gaps | ||
| Pixel size | 13.5 microns square; 0.0727 arcsec/pixel | ||
| Spectral Response | approx 0.3 to 0.9 microns [ data / plot ] | ||
| Bias level | Bias image | ||
| Flat field response | Fringe images Flat field images | ||
| Readout time | See observing overheads page | ||
| Chip | CCD 01 | CCD 02 | CCD 03 |
| Chip ref no. | EEV 9273-16-03 | EEV 9273-20-04 | EEV 9273-20-03 |
| Dark current* | 0.8 e-/pix/hr | 0.7 e-/pix/hr | 0.5 e-/pix/hr |
| Full Well ** | 150 ke- | 101 ke- | 159 ke- |
| Fringing at 900nm *** | 29% | 19% | 24% |
* Dark current measured at -120C by NOAO
** Saturation can be avoided by assuming 100ke- for all chips
*** (peak-valley)/mean, measured from images taken at NOAO
Readnoise and Gain Values
The table below gives gain/read-noise values for the current GMOS EEV CCDs operating with the SDSU controller. Only the suggested settings are listed. The values are averaged over all 6 amps. For the full table of read-noise and gain values at all settings and for all amps, click here.
| Readout | Gain | Resulting average | |
| rate | level | Gain (e-/DN) | noise (e- rms) |
| Fast | High | 5.02 | 7.4 |
| " | Low | 2.49 | 4.9 |
| Slow | High | 4.40 | 4.8 |
| " | Low | 2.18 | 3.4 |